A kinetic model for the thermoluminescent high dose response of LiF:Mg, Cu,P (MCP-N)
Loading...
Files
Identifiers
Publication date
Abstract
This contribution describes a kinetic model attempting to reproduce the response of the thermoluminescent
material LiF:Mg,Cu,P when it is irradiated to absorbed dose values in the kGy range. The modelling is based on
the hypothesis of a relationship between the irradiation time (i.e. the absorbed dose) and the density of trapping/
recombination centres.
X-ray diffraction and thermal X-ray diffraction measurements have been performed to investigate the potential
radiation and thermal damage on the structure of the material, including the possibility of partial phases. The
proposed kinetic model qualitatively reproduces the observed changes in the TL glow curve for temperatures
above the main peak as well as the two observed regions of absorbed dose response: linear and sub-linear.