Título : Monte Carlo modelling for the in vivo lung monitoring of enriched uranium: Results of an international comparison
Autor : Pérez López, Begoña
Navarro Amaro, Juan Francisco
López Ponte, María Antonia
Palabras clave : Montecarlo
calibration
internal dosimetry
intercomparison
Fecha de publicación : 29-dic-2023
Resumen : In order to assess the reliability of Monte Carlo (MC)-based numerical calibration of in vivo counting systems the EURADOS network supported a comparison of MC simulation of well-defined experiments. This action also provided training for the use of voxel phantoms. In vivo measurements of enriched uranium in a thoracic phantom have been carried out and the needed information to simulate these measurements was distributed to 17 participants. About half of the participants managed to simulate the measured counting efficiency without support from the organisers. Following additional support all participants managed to simulate the counting efficiencies within a typical agreement of ±5% with experiment.