Comparison of two leg phantoms containing 241Am in bone.

dc.contributor.authorPérez López, Begoña
dc.contributor.authorLópez Ponte, María Antonia
dc.contributor.authorNavarro Amaro, Juan Francisco
dc.date.accessioned2023-12-29T18:31:44Z
dc.date.available2023-12-29T18:31:44Z
dc.date.issued2023-12-29
dc.description.abstractThree facilities (CIEMAT, HMGU and HML) have used their in vivo counters to compare two leg phantoms. One was commercially produced with (241)Am activity artificially added to the bone inserts. The other, the United States Transuranium and Uranium Registries' (USTUR) leg phantom, was manufactured from (241)Am-contaminated bones resulting from an intake. The comparison of the two types of leg phantoms showed that the two phantoms are not similar in their activity distributions. An error in a bone activity estimate could be quite large if the commercial leg phantom is used to estimate what is contained in the USTUR leg phantom and, consequently, a real person. As the latter phantom was created as a result of a real contamination, it is deemed to be the more representative of what would actually happen if a person were internally contaminated with (241)Am.es_ES
dc.identifier.doihttp://dx.doi.org/10.1097/HP.0b013e3182118f61
dc.identifier.urihttps://hdl.handle.net/20.500.14855/2107
dc.language.isoenges_ES
dc.rights.accessRightsopen accesses_ES
dc.subjectAmericiumes_ES
dc.subjectknee phantomes_ES
dc.subjectin vivoes_ES
dc.titleComparison of two leg phantoms containing 241Am in bone.es_ES
dc.typejournal articlees_ES

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
11.HP_Comparison leg phantoms.pdf
Size:
2.58 MB
Format:
Adobe Portable Document Format